Twenty small data loggers sit in a cardboard tray on the QA desk, each one with a strip of masking tape and a code written in marker: A1-low, A1-mid, A1-high. The engineer has a floor plan of the cold room with pencil crosses on it. By Friday the loggers will be taped to racks, and by Monday nobody will remember why the crosses went where they did.
Cold room temperature mapping is a documented study that places calibrated loggers across the room, at several heights, for a set period, to find the hot and cold spots and prove the room holds its range. Its real output is not the report. It is a decision about where the routine probes of your cold room temperature monitoring go, and what the alarm limits on them should be.
Mapping proves the whole room holds the range, not the controller probe
The controller probe sees one spot. Product sits everywhere else: near the door seal, under the evaporator, on the top shelf where warm air pools. The WHO technical supplement on temperature mapping of storage areas (TRS 961, Annex 9, Supplement 8) says a mapping study establishes the temperature distribution in a space and locates its hot and cold spots. It applies to any space holding product with a labelled storage temperature, including cold rooms, freezer rooms, quarantine areas and loading bays.
In India, CDSCO's draft Guidelines on Good Distribution Practices for Pharmaceutical Products say storage areas shall be temperature mapped under representative conditions, and recommend that monitors sit where fluctuations are most likely. That second clause is the one people skip. Your QA and validation advisers decide what applies to your licence; the logic of both documents is the same.
The study has four stages, and the last one is the point
WHO describes four stages: prepare a protocol, carry out the mapping, write a report, then implement its recommendations. The fourth stage is where most studies stall. The report gets signed and filed, and the routine probes stay where the cable was easiest to run.
| Decision | What the WHO supplement says | What you still decide |
|---|---|---|
| Grid across the floor | Loggers in a grid along length and width, typically every 5 to 10 metres | Extra points for alcoves, racking and where product really sits |
| Heights | Rooms up to 3.6 m high: low, middle and high, for example floor, 1.2 m and 3.0 m | Whether top shelves near the evaporator need their own point |
| Recording interval | The same interval on every logger, typically 1 to 15 minutes, same start time | One interval that suits the room's defrost and door pattern |
| Duration | Cold rooms and freezer rooms: 24 to 72 hours, or longer if justified | Whether to include planned door openings or a power cut |
| Load state | New rooms mapped empty and in a normal loaded condition | What 'normal load' means for your busiest week |
| Remapping | Periodically, for example every three years, and after significant changes such as a set point change | Your own trigger list, written into the SOP |
If the room has two refrigeration units, the supplement says to map with each unit running on its own. The temperature pattern can change with whichever unit is carrying the load.
A worked example: placing loggers in a 15 by 10 metre room
The numbers below are illustrative, not a rule and not from any site. Say a walk-in cold room is 15 m long, 10 m wide and 3.4 m high, held at 2 to 8 °C.
- Grid at 5 m spacing: positions at 2.5, 7.5 and 12.5 m along the length, and 2.5 and 7.5 m across the width. That is 3 × 2 = 6 grid points.
- The ceiling is under 3.6 m, so three heights per point: floor, 1.2 m and 3.0 m. That is 6 × 3 = 18 loggers.
- Add one beside the door and one under the evaporator, where risk is obvious: 18 + 2 = 20 loggers.
- At a 5-minute interval over 72 hours, each logger records 72 × 60 ÷ 5 = 864 readings, so the study holds 20 × 864 = 17,280 readings in all.
Now say the results show a hot spot of 7.4 °C at the top level beside the door after loading, and a cold spot of 2.6 °C on the low shelf under the evaporator. Both are inside 2 to 8 °C, so the room passes. But the margins are 8 − 7.4 = 0.6 °C at the top and 2.6 − 2 = 0.6 °C at the bottom. That is where two of your routine probes belong, and those shelves may suit less sensitive stock.
A mapping report that does not move a single probe has measured the room and changed nothing.
The hot and cold spots decide your probes and your alarm limits
The supplement is direct about this: hot and cold spots are found to identify where monitoring sensors should go. It also lists repositioning monitoring and control sensors among the outcomes of a report. So the routine probes follow the map, and the alarm limits follow the probes.
A probe at the hot spot needs an alert limit low enough to fire before the label limit is reached there. A probe at the cold spot needs a low alarm that matters, since many refrigerated products are harmed by freezing. Our post on what temperature a pharma cold room should be walks through set points, alert limits and action limits. Put those numbers next to your map and check that every critical spot has a probe with a limit that leaves time to act.
The probe you add for the door should not alarm on every loading cycle. WHO allows a protocol to define door openings and accept a return to range within 30 minutes. Use the study to learn how long your room takes to recover, then set the alarm delay from that, not from a guess.
Mean kinetic temperature belongs after the map, not inside it
The mapping supplement analyses minimum and maximum temperatures and arithmetic means per logger location. Mean kinetic temperature (MKT) answers a different question: how much thermal stress did product at one spot see over a period? It weights warm readings more than a plain average, because degradation speeds up with heat.
Take an illustrative location that reads 5.0 °C for 18 hourly readings and 8.0 °C for 6. The arithmetic mean is (18 × 5 + 6 × 8) ÷ 24 = 5.75 °C. The MKT, at the common 83.144 kJ/mol default, works out to about 5.86 °C. The gap is small here and grows with warmer peaks. Run your own hot-spot data through the mean kinetic temperature calculator to see it, and treat the result as an input to QA's judgement during an excursion review, not as a mapping pass mark.
"Our room is small and the display is always 5 °C"
This is a fair objection. A small reach-in unit with good airflow may show very little spread, and a full grid feels like ceremony. The supplement itself allows a risk-based approach to logger locations, and a small space needs fewer of them.
But the display shows the controller probe, which is placed to control the unit, not to find its worst corner. The only way to know the spread is small is to measure it once, properly. A short, honest study of a small room is cheap. Discovering a cold shelf during an inspection, with vaccines on it, is not.
A map goes stale when the room changes
New racking, a heavier loading pattern, a replaced evaporator or a set point change all alter the airflow. WHO lists these as reasons to remap, along with unexplained variability in monitoring records. Continuous data helps here: if a probe at a former cold spot starts drifting, that is a prompt to look again. Systems such as ColdWatch can read existing sensors over BACnet/IP or Modbus TCP and point to likely causes, including sensor placement, but the decision to remap stays with QA. For how a drift becomes an investigation, read our guide to investigating a cold room temperature excursion.
What to do this week
- Find your last mapping report and note its date, the load state and the season it covered.
- Mark its hot and cold spots on today's floor plan, and check whether a routine probe sits at each one.
- Compare each probe's alert limit with the margin the map found at that spot.
- List what has changed since the study: racking, equipment, set point, loading pattern.
- If the list is long, ask QA whether a remap belongs in this quarter's plan.
The tray of taped loggers is the easy part. The value is in the line on the report that says "move probe 3 to the top shelf by the door", and in someone actually moving it. More on monitoring and alarms in our cold chain monitoring guides.